Ureteral stents have become an integral part of tulas. Straight silicone stents provided good
contemporary urologic practice. They are typ-
internal drainage and developed less encrustation
ically placed to prevent or relieve ureteral obstruc-
than other compounds, but had no distal or prox-
tion due to a variety of intrinsic or extrinsic etiol-
imal features to prevent migration. Minor im-
ogies. These include obstructing ureteral calculi,
provements followed, with closing the proximal
ureteral strictures, congenital anomalies such as
end of the stent to facilitate passage and using a
ureteropelvic junction obstruction, retroperito-
“pusher” to hold the stent in place during wire
neal tumor or fibrosis, trauma, and iatrogenic in-
jury. Stents are also placed to provide urinary di-
modifications to prevent stent migration, includ-
version or postoperative drainage or to help
ing a distal flange to prevent proximal migration
identify and prevent inadvertent injury to the ure-
and sharply pointed barbs to prevent downward
migration and expulsion. In 1974, the Gibbonsstent became the first commercially available
HISTORY OF STENTS
“modern” internal ureteral stent. Following the
The concept of stenting the urinary system began
Gibbons stent, single-pigtail stents were designed
as an adjunct to open surgery to facilitate upper
that could be straightened and placed over a wire
tract drainage or to align the ureter. Gustav Simon
cystoscopically to prevent distal, but not proximal,
performed the first reported case during the 19th
century by placing a tube in the ureter during open
solved with the use of double-J stents, which con-
cystostomy, and Joaquin Albarrano created the
first catheter intended for use in the ureter in the
The word “stent” is commonly used in genitouri-
nary reconstructive surgery. Two related uses of
was first reported in 1839, early catheters were
the stent are given in most medical dictionaries.
First, a stent is a device used to maintain a bodily
development of plastics, such as polyethylene and
orifice, cavity, or contour, and second, it is a cath-
polyvinyl, allowed for stents to become more rigid
eter, rod, or tube within a tubular structure to
and easier to place; however, problems such as
maintain luminal patency or protect an anastomo-
bladder irritation, infection, encrustation, and mi-
sis or Etymologically, the word “stent” is an
eponym related to three English dentists named
using polyethylene tubes to help repair ureters and
Stent who contributed to improving a substance
In 1967, the era of the modern long-term in-
dental impression compound into a urologic tool is
dwelling ureteral stent began when Zimskind and
attributed to of Holland during World War
I, as the name became associated with a support for
cone tubing inserted endoscopically to bypass ma-
oral skin grafts and later adopted in plastic surgery.
lignant ureteral obstruction or ureterovaginal fis-
Urologic stenting first appeared in print beginningin the 1970s. wrote a brief commentaryin 1972 titled, “Splint, Stent, Stint,” concluding:
From the Department of Urology, University of California, David
“Urologists are always talking about putting a tube
Geffen School of Medicine, Los Angeles, Los Angeles, California;
in a ureter or urethra. When they do this, it is not a
and Department of Urology, New York-Presbyterian Hospital,Columbia University College of Physicians and Surgeons, New
splint. It may be a stent. It probably is never a stint.
Perhaps the process is most properly described as
Reprint requests: Mantu Gupta, M.D., Department of Urology,
leaving a tube or stent in an organ.” The word
Columbia University College of Physicians and Surgeons, 161
“stent” became forged into the urologic literature
Fort Washington Avenue, Irving Pavilion, 11th Floor, New York,NY 10032
when Montie et explicitly defined the term:
Submitted: December 29, 2003, accepted: March 1, 2004
“When referring to an intraluminal device to main-
UROLOGY 64: 9 –15, 2004
tain patency until healing has taken place, the stent
teral replacement using glass, tantalum, or vital-lium were first described before World War In
PROPERTIES OF STENT MATERIALS
contemporary practice, devices constructed from
The modern self-retained internal ureteral stent
synthetic polymeric compounds are the most use-
is a synthetic polymeric biomaterial device that can
ful materials implanted in the urinary tract. Cur-
be placed by a variety of endourologic techniques.
rently, self-retained internal ureteral stents are
It is designed to retain its position within the ureter
manufactured from “high polymers.” A polymeric
and drain urine between the renal pelvis and blad-
chain has tens of thousands of chemical “mono-
der. The basic properties of a ureteral stent have
mers,” which are linked by “polymerization.”
been reviewed by Mardis et “Memory” pro-
These chains are organized in various configura-
duces the ability to maintain the position within
tions resulting in complex and versatile macromol-
the ureter and is a consequence of polymeric cross-
linking, which results from physical or chemical
Polyethylene was the first synthetic polymer
bonding between macromolecular chains. “Du-
used to fashion ureteral stents. It consists of poly-
rometer” refers to the strength of the memory,
olefin and is flexible, odorless, translucent, and
which can be varied within the same material and
nonreactive in the body. The stiffness of polyeth-
depends on the type of cross-linking used. As
ylene made it useful for the management of ure-
cross-linking increases, the durometer changesfrom soft to hard. Elasticity allows the shape of the
teral strictures, but it was found to promote protein
stent to be manipulated. Variations in tensile
deposits, leading to an increased likelihood of crys-
strength reflect the mechanism of crystallization
and cross-linking in the biomaterials. Greater ten-
Furthermore, long-term exposure of polyethylene
sile strength allows for the creation of a higher
internal/external diameter ratio and a greater num-
composed of alternating silicone and oxygen at-
It is nonirritating and resistant to encrusta-
elongation capacity is the percentage of elongation
tion. However, silicone stents migrate easily, have
at stent breakage. Thermoplastic elastomeres, such
poor mechanical strength, and have a high coeffi-
as polyurethanes and a variety of other proprietary
cient of friction, making it difficult to use when
copolymers, generally have more elongation ca-
negotiating strictures or tortuous ureters. Polyure-
pacity than thermoset elastomeres, such as sili-
thane, a common generic class of condensation
Biodurability refers to the ability to exist
polymers derived from polyisocynate and a polyol
within the body without degradation of stent struc-
has been used in an effort to combine the flexibility
ture and function. The urinary tract presents an
of silicone with the stiffness of polyethylene. Poly-
unstable and hostile environment to stent func-
urethane is highly versatile and inexpensive, but
tion, replete with free radicals, oxidizers, enzymes,
has been shown to induce more epithelial ulcer-
and supersaturation of urinary mucoids and crys-
talloids. Stent degradation within the urinary tract
ited durability and demonstrates slow in vivo bio-
has been reported for polyethylene, polyurethane,
and silicone internal ureteral Biocompat-
Proprietary modifications to silicone led to the
ibility is the utopian state at which the material
development of C-Flex (Consolidated Polymer
present within the ureter has no significant effect
Technologies, Clearwater, Fla), a silicone-modi-
on the interface between itself and the urothelium,such as the ability to resist encrustation and infec-
fied styrene/ethylene/butylene block thermoplastic
tion. The biocompatibility of a stent may be en-
copolymer. The partial silicone composition of C-
hanced with use of hydrophilic polymers that have
Flex makes these stents softer than polyurethane
low protein absorption and low bacterial adher-
and theoretically less likely to develop encrusta-
The coefficient of friction describes how
tion. Silitek (ACMI, Southborough, Mass) is a pro-
easily a stent is passed or exchanged. Surface fric-
prietary, second-generation, polyester copolymer
tion is generally dependent on the durometer and a
that followed silicone and polyurethane. Percuflex
greater durometer lowers the coefficient of friction.
(Boston Scientific, Natick, Mass) is a proprietary
Applying a hydrophilic coating reduces the surface
olefinic block copolymer developed by Boston Sci-
coefficient of friction. Radiopacity refers to the ease
entific/Microvasive that becomes soft and flexible
of stent visualization during fluoroscopy. All stents
at body temperature. This material also has excel-
are radiopaque and some contain fillers that fur-
lent memory and strength. Tecoflex (Thermedics,
Woburn, Mass) is a proprietary aliphatic thermo-plastic polymer that is not silicone based. Stents
STENT BIOMATERIALS AND COATINGS
made with this material have a smooth surface and
The use of biomaterials within the urinary tract
a comparatively large inside diameter. This mate-
dates back to ancient Egypt when lead and papyrus
rial allows for ink labeling instead of laser labeling,
UROLOGY 64 (1), 2004
which may theoretically decrease the likelihood of
tients with strictures of various etiologies and lo-
encrustation from the rough surface created by the
laser. Sof-Flex (Cook Urological, Spencer, Ind) is a
be placed either alone or in combination with con-
proprietary compound from Cook Urological.
ventional indwelling ureteral stents in patients
Other stent materials from Cook include Endo-Sof
with long strictures. These stents gradually be-
and Ultrathane. Pellethane (C.R. Bard, Covington,
come covered with urothelium after implantation,
Ga) is a proprietary copolymer from Bard used in
thus preventing encrustation and infection. The
their Inlay stents. Flexima is a newer proprietary
problems associated with metal stents include col-
material from Boston Scientific/Microvasive de-
lagenous ingrowth, hyperplastic epithelium, distal
signed to resist buckling during insertion. Vertex
ureteral narrowing, intense fibrosis, and subse-
(Applied Medical, Rancho Santa Margarita, Calif)
is a proprietary material from Applied Medical.
The development of tissue-engineered stents is
Bioabsorbable polymeric materials are designed
in its infancy, but this area has the potential to
to retain their tissue-supporting properties for de-
revolutionize biomaterial science and the clinical
fined periods. After placement, they are gradually
applications of biomaterials. The use of autologous
biodegraded into tissue-compatible compounds
tissue would be advantageous because of its inher-
that are absorbed and replaced by healing tissue.
ent biocompatibility. Cartilaginous stents have
These stents will be most useful for clinical situa-
been successfully created in vitro and in vivo using
tions in which temporary upper urinary tract
drainage is desired. Bioabsorbable materials used
technology might one day become clinically useful
in urologic stents are high-molecular-weight poly-
in the treatment of urethral or ureteral strictures.
mers of polylactic and polyglycolic Poly-L-
The goals of coating ureteral stents are to facili-
lactide-co-glycolide bioabsorbable ureteral stents
tate passage over a guidewire beyond an obstruc-
have been shown to be similar to standard ureteral
tion and to reduce or eliminate biofilm formation
stents with respect to fluid flow results and post-
and encrustation. Hydrophilic coatings consist of
operative appearance on retrograde pyelography in
nondissolvable polymers that swell on contact
a porcine model after cutting balloon device endo-
with water and retain water within their polyan-
ionic structure that layers on its surface. The sur-
have less favorable biocompatibility than standard
face water not only reduces the coefficient of fric-
stents because of inferior healing of the incised
tion, improving ease of stent insertion, but also
ureteral musculature. Temporary ureteral drainage
contributes to biocompatibility by reducing fric-
stents (Boston Scientific/Microvasive) have been
tional irritation and cell adhesion at the biomateri-
al-urothelial interface. LSe is Cook Urological’s ion
more, dissolution of these stents can be monitored
implantation process in which specific ions are im-
by radiographic examination during a period of 2
planted into the chemical structure. LSe is de-
to 7 days owing to the inclusion of a radiopaque
signed to reduce the coefficient of friction without
filling in the bulk biomaterial. In a multicenter
presoaking in water and lower the surface energy
Phase II clinical trial, 90% of the stents dissolved as
of the AQ is a hydrophilic coating made by
expected with a median elimination time of 8 days,
Cook Urological that can be applied to lower the
and 89% of the patients were satisfied with their
coefficient of friction. Other hydrophilic coatings
include SL-6 by Applied Medical and Hydro Plus
pand at body temperature to hold their position
by Boston Scientific/Microvasive. A new hydro-
philic coating, polyvinylpyrrolidone, has been
stents do not communicate with the bladder or
shown in vitro to decrease both hydroxyapatite en-
kidney and would avoid vesicoureteral reflux and
crustation and adherence of a hydrophobic Entero-
possibly would have a lower infection rate. An al-
coccus faecalis Another coating that has
ternative strategy is o control degradation by mod-
potential for reducing encrustation is the phospho-
ulating the urinary pH to an alkaline level when the
rylcholine group of compounds. Phosphorylcho-
line-coated stents have been shown to be less vul-
Metallic, superalloy titanium and nickel/tita-
nerable to encrustation and colonization by
nium are materials currently used for permanently
bacterial biofilm than the uncoated stents at 12
implanted stents, such as the Wallstent (Medin-
rent, Lausanne, Switzerland) or the Memokath 051
studies, heparin-like polysaccharides appear to
(Engineers and Doctors of Copenhagen, Copenha-
have great potential to reduce biomaterial encrus-
gen, Denmark), to relieve conditions such as ma-
tation. Heparin-coated stents have been shown not
to have any biofilm formation or encrustation after
Varying success rates using these devices have
been reported, ranging from 14% to 100%, in pa-
rial-related urinary tract encrustation involves
UROLOGY 64 (1), 2004
coating biomaterials with oxalate-degrading en-
The Towers peripheral stent is a grooved stent
zymes derived from Oxalobacter formigenesA re-
manufactured by Cook Urological. The outer di-
duction in the amount of encrustation on oxalyl
ameter ranges from 6 to 8 F and the length from 22
coenzyme A decarboxylase-coated and formyl co-
to 32 cm. The LithoStent by ACMI features three
enzyme A transferase-coated disks versus control
grooves spiraling down the exterior of stent-like
disks was demonstrated in an in vivo rabbit blad-
gutters designed to prevent ureteral strictures and
der implantation model after a 30-day implanta-
A recent study has shown that silver nitrate
Spiral stents are polyurethane stents supported
and ofloxacin-blended caprolactam-L-lactide co-
by a built-in metal spiral wire intended to keep the
polymer-coated self-reinforced poly-L-lactic acid
stent patent and improve drainage. In a study of 14
Urospiral urethral stents have good biocompatibil-
patients with spiral stents placed for chronic ure-
ity and caused less tissue reaction than stents with
teral obstruction, only one stent change occurred
a pure copolymer coating in a rabbit model. Fur-
thermore, the silver nitrate coating was effective in
spiral-ridged stents were compared with smooth-
preventing biofilm formation and stent encrusta-
walled double-J stents in an in vitro mechanical
ureteral model, extraluminal flow was greatestwith the 7F spiral-ridged
STENT DESIGNS
A mesh stent is a lightweight, self-expanding
stent designed to preserve drainage while minimiz-
A double-J stent has an open hook configuration
ing irritative symptoms. In a porcine model, al-
at either end of the ureteral stent, and a double-
though not statistically significant, the mesh stent
pigtail stent has a full retentive coil at either end.
resulted in less inflammation along the urinary
The double-J stent is the original, silicone, closed-
tract at 1 week than a standard 7F double-pigtail
tipped stent to which ACMI/Surgitek holds the
through the mesh stent tended to be greater than
double-pigtail stent, which have a variety of names,
the flow through the standard stent at both 1 and 6
designs, and compositions. Standard double-pig-
weeks. When the flow characteristics were com-
tail stents must be sized correctly because those
pared with other stents in a separate porcine ani-
that are too long can cause bladder irritation, and
mal study, the mesh stent had the greatest fl
those that are too short can migrate up the ureter. A
Tail stents have a standard pigtail and a 6F or 7F
24-cm ureteral stent is well suited for most adults,
shaft at the proximal end, but the distal end tapers
but should be individualized according to the ure-
into an elongated 3F closed-tip tail to decrease
teral length of the patient. Multiple-coil stents weredesigned to meet the “one size fits all” concept.
stent-related bladder irritability. Drainage is
Multiple-coil stents leave redundant length in the
achieved around the distal portion of tail stents,
bladder coiled, so the trigone is not irritated. Some
and, consequently, they are contraindicated when
stents are manufactured with a nylon suture at-
trauma to the distal ureter is suspected. Tail stents
tached to the distal end, which can be left at the
resist reflux because the distal third of the tail is
urethral meatus, allowing for removal after short-
occluded and approximately 3 to 5 cm of the tail’s
term drainage without the need for repeat cystos-
occluded portion lies within the intramural tunnel
and the distal ureter. Boston Scientific/Microvasive
Open-ended ureteral stents have no coils and
makes a tail stent called the Percuflex Tail Plus.
thus are not suitable for long-term urinary drain-
Tail stents have been shown to produce signifi-
age. However, they can be secured externally to
cantly fewer irritative voiding symptoms and result
provide temporary drainage. Open-ended stents
in significantly less postoperative flank pain after
are useful in helping to direct and advance a guide-
wire into a ureteral orifice. They can assist in col-
Dual-durometer stents incorporate a smooth
lecting upper urinary tract urine samples and per-
transition from a firm biomaterial at the renal end
mitting retrograde pyelography. Furthermore,
to a soft biomaterial at the bladder end. Two such
open-ended ureteral stents are commonly placed
devices are the Sof-Curl (ACMI) and Polaris (Bos-
intraoperatively to identify ureters to avoid inad-
ton Scientific/Microvasive) stents, which have long
vertent injury during abdominal or pelvic surgery.
tapered tips at the renal end and are coated with a
Transilluminating ureteral stents have been used
hydrophilic-bonded hydrogel that decreases their
to prevent ureteral injuries during gynecologic
ureteral stents, the Polaris stent was shown to have
Grooved stents have grooves spiraling down the
superior lubricity, which eases stent passage, and
exterior of the stent to improve extraluminal flow
lower flexural strength, which minimizes bladder
aimed at postlithotripsy or holmium laser cases.
UROLOGY 64 (1), 2004
Magnetic material-tipped ureteral stents were
tion, in whom indwelling ureteral stents have
developed to eliminate the morbidity of cysto-
failed. The proximal end of the stent is inserted
scopic stent removal. The use of magnetic stents
into the renal pelvis by way of a percutaneous ne-
was initially reported in 1989, when Macaluso and
phrostomy puncture. A subcutaneous tunnel is
created from the flank to the bladder, and the distal
tients. Retrieval of the Magnetip (ACMI/Surgitek)
end of the stent is passed into the bladder through
double-J stent was accomplished in 86% of cases
using the Magnetriever (ACMI/Surgitek). Mag-
Fistula stent sets specifically assist in the man-
netic stents have also been used in an attempt to
agement of ureteral fistulas. The stent has drainage
eliminate the need for general anesthesia or cystos-
ports solely in the pigtails to reduce fluid pressure
copy in children. In 7 children, the stents were
within the ureter. It is manufactured in several dif-
removed with a magnetic-tip retriever without the
need for cystoscopy, and only 1 patient requiredMore recently, an approach that uses
GUIDEWIRES
a powerful rare-earth magnet on the retrieval cath-
Accurate and atraumatic stent placement is
eter and a stainless steel bead attached to the end of
greatly enhanced by proper ureteral access using a
the stent has been developed. The magnetic stent
guidewire under fluoroscopic guidance. Guide-
was removed without difficulty in 29 (97%) of 30
wires straighten the course to the ureter and secure
patients, and no complications related to the place-
the luminal path for passage of a stent, balloon
dilator, or ureteroscope. Different types of guide-
Injection stents are single-pigtail catheters that
wires have been developed; each designed with
convert into an indwelling double-pigtail ureteral
unique structural properties to help accomplish
stent, obviating a second cystoscopy. It is ideal for
specific tasks. Guidewires are described by mate-
patients undergoing shockwave lithotripsy who
rial of composition, size (diameter and length), tip
need a double-pigtail stent postoperatively, but si-
design, surface coating, core (moveable or non-
multaneously require contrast instillation during
movable), and shaft rigidity. Guidewires may be
the procedure for stone localization. Once the pro-
composed of stainless steel or nitinol. Stainless
cedure is completed, the injection apparatus is re-
steel wires may be round or flat. Round wires have
moved, leaving a double-pigtail stent.
a standard stiffness, and flat wires allow for a
Urinary diversion stents have a single coil for the
thicker, stiffer core without an increased diameter.
renal pelvis and a long straight end that is brought
Nitinol-based wires are kink resistant and can be
out for external drainage. The drainage holes are in
designed with a stiffer core or a stripped core. Stan-
the proximal end of the stent. They are commonly
dard guidewires range in diameter from 0.018 to
used to provide internal support to ureteral anas-
0.038 in., and available lengths vary from 145 to
tomoses created during various types of urinary
180 cm. The distal 3 cm of guidewires are designed
diversion. This allows for accurate monitoring of
to be soft and flexible to minimize trauma to the
urine output from each renal unit and facilitates
ureteral wall during passage. Various tip designs
removal of the stent after an adequate period of
and shapes have been developed to facilitate pas-
sage through tortuous ureteral anatomy or ureteral
strictures. Tip configurations may be straight, an-
used for temporary drainage from the ureteropel-
gled, or J-shaped. Straight and angled guidewires
vic junction to the bladder after incision of a stric-
are also available with an 8-cm flexible tip. Shaft
ture. These internal stents have a tapering diameter
rigidity depends on the core materials. Guidewires
with no sideports and are designed for postendo-
may be coated with polytetrafluoroethylene
pyelotomy procedures to prevent narrowing of the
(PTFE, Teflon), which improves advancement of
ureteral lumen while preventing ingrowth of the
more rigid devices; a lubricious covering, used for
ureteral wall to the stent. The larger caliber portion
soft devices; or a hydrogel, which permits near fric-
of the stent acts as a mold for the healing of the
incised ureter. Applied Medical has a 7/10F endo-pyelotomy stent for use with the Acucise endopy-
FUTURE DIRECTIONS
elotomy balloon catheter device. Cook Urologicalmakes a 6/10F and 7/14F stent, and Boston Scien-
Advances in surface-coating technologies, along
tific/Microvasive makes the RetroMax Plus in
with the use of biodegradable materials will result
in novel drug-delivery systems. By incorporating
Subcutaneous urinary diversion (nephrovesical)
various drugs and medications, such as growth fac-
stents are used as an alternative to nephrostomy for
tors, anti-inflammatory agents, and chemothera-
urinary diversion in uremic cancer patients and
peutic drugs, new therapeutic strategies for the
patients with benign idiopathic ureteral obstruc-
treatment of ureteral stricture disease and, possi-
UROLOGY 64 (1), 2004
bly, genitourinary malignancy may be offered.
9. Hepperlen TW, Mardis HK, and Kammandel H: Self-
Drugs could be applied directly to the polymeric
retained internal ureteral stents: a new approach. J Urol 119:
surface of the biomaterial and be pharmacologi-
10. Finney RP: Experience with new double J ureteral
cally active at the surface. Alternatively, drugs
catheter stent. J Urol 120: 678 –681, 1978.
could be incorporated into the core of the poly-
11. Bloom DA, Clayman RV, and McDougal E: Stents and
meric structure of the biomaterial and delivered to
related terms: a brief history. Urology 54: 767–771, 1999.
the local environment in a sustained-release fash-
12. Esser JF: Studies in plastic surgery of the face. Ann Surg
ion and, consequently, have potential activity at
65: 297–315, 1917.
13. Goodwin WE: Splint, stent, stint. Urol Dig 11: 13–14,
the level of the urothelium. An interesting ap-
proach to achieve controlled release of a drug in
14. Montie JE, Stewart BH, and Levin HS: Intravasal stents
the ureter has used multilamellar liposomes en-
for vasovasostomy in canine subjects. Fertil Steril 24: 877–
trapping dexamethasone. In an in vitro setting,
51% of the drug was released during a 48-hour
15. Mardis HK, Kroeger RM, Morton JJ, et al: Comparative
evaluation of materials used for internal ureteral stents. J En-
dourol 7: 105–115, 1993.
tinue to be developed to decrease the morbidity
16. Mardis HK, Kroeger RM, Hepperlen TW, et al: Polyeth-
associated with indwelling stents. Advances in
ylene double-pigtail ureteral stents. Urol Clin North Am 9:
other subspecialties, such as endoluminal paving
in cardiac stenting, may offer insights into segmen-
17. Denstedt JD, Wollin TA, and Reid G: Biomaterials used
in urology: current issues of biocompatibility, infection, and
encrustation. J Endourol 12: 493–500, 1998.
approach for treating ureteral or ureterointestinal
18. Bitschay J, and Brodny ML: A History of Urology in
strictures. Finally, research into ureteral replace-
Egypt. New York, Riverside Press, 1956, p 76.
ment techniques may redefine the role of ureteral
19. Lord JW, and Eckel JH: The use of vitallium tubes in the
urinary tract of dogs. J Urol 48: 412–420, 1942.
20. Cormio L, Talja M, Koivusalo A, et al: Biocompatibility
CONCLUSIONS
of various indwelling double-J stents. J Urol 153: 494 –496, 1995.
Indwelling ureteral stents offer urologists an
21. Brauers A, Jung PK, Thissen H, et al: Biocompatibility,
enormous arsenal against a host of urologic dis-
cell adhesion, and degradation of surface-modified biodegrad-able polymers designed for the upper urinary tract. Tech Urol
eases. Ongoing research is essential to optimize
4: 214 –220, 1998.
biocompatibility and decrease stent-related com-
22. Olweny EO, Landman J, Andreoni C, et al: Evaluation
plications. Interest has been great in this field of
of the use of a biodegradable ureteral stent after retrograde
research, marked by the introduction of new bio-
endopyelotomy in a porcine model. J Urol 167: 2198 –2202,
materials, coatings, and stent designs. An ideal
23. Auge BK, Ferraro RF, Madenjian AR, et al: Evaluation
stent does not exist, so it is important that urolo-
of a dissolvable ureteral drainage stent in a swine model. J Urol
gists be familiar with the various indications for
168: 808 –812, 2002.
use, selection, modes of insertion, and potential for
24. Lingeman JE, Preminger GM, Berger Y, et al: Use of a
complications. With this information, urologists
temporary ureteral drainage stent after uncomplicated uret-
will optimize the efficacy and safety of ureteral
eroscopy: results from a phase II clinical trial. J Urol 169: 1682–1688, 2003.
25. Lumiaho J, Heino A, Tunninen V, et al: New bioabsorb-
able polylactide ureteral stent in the treatment of ureteral le-
sions: an experimental study. J Endourol 13: 107–112, 1999.
1. Herman JR: Urology: A View Through the Retrospectro-
26. Schlick RW, and Planz K: In vitro results with special
scope. Hagerstown, Maryland, Harper & Row, 1973.
plastics for biodegradable endoureteral stents. J Endourol 12:
2. Mack T: The double-J: then and now, in Yachia D (Ed):
Stenting the Urinary System. Oxford, ISIS Medical Media,
27. Pauer W, and Eckerstorfer GM: Use of self-expanding
permanent endoluminal stents for benign ureteral strictures:
3. Herdman JP: Polyethylene tubing in the experimental
mid-term results. J Urol 162: 319 –322, 1999.
surgery of the ureter. Br J Surg 37: 105–106, 1949.
28. Kulkarni R, and Bellamy E: Nickel-titanium shape
4. Tulloch WS: Restoration of continuity of the ureter by
memory alloy Memokath 051 ureteral stent for managing
means of a polyethylene tubing. Br J Urol 24: 42–45, 1952.
long-term ureteral obstruction: 4-year experience. J Urol 166:
5. Zimskind PD, Kelter TR, and Wilkerson SL: Clinical
use of long-term indwelling silicone rubber ureteral splints
29. Amiel GE, Yoo JJ, Kim BS, et al: Tissue engineered
inserted cystoscopically. J Urol 97: 840 –844, 1967.
stents created from chondrocytes. J Urol 165: 2091–2095,
6. Marmar JL: The management of ureteral obstruction
with silicone rubber splint catheters. J Urol 104: 386 –389,
30. Denstedt JD, Reid G, and Sofer M: Advances in ureteral
stent technology. World J Urol 18: 237–242, 2000.
7. Orikasa S, Tsuji I, Siba T, et al: A new technique for
31. Tunney MM, and Gorman SP: Evaluation of a poly(vi-
transurethral insertion of a silicone rubber tube into an ob-
nyl pyrrolidone)-coated biomaterial for urological use. Bioma-
structed ureter. J Urol 110: 184 –187, 1973.
terials 23: 4601–4608, 2002.
8. Gibbons RP, Mason JT, and Correa RJ Jr: Experience
32. Stickler DJ, Evans A, Morris N, et al: Strategies for the
with indwelling silicone rubber ureteral catheters. J Urol 111:
control of catheter encrustation. Int J Antimicrob Agents 19:
UROLOGY 64 (1), 2004
33. Riedl CR, Witkowski M, Plas E, et al: Heparin coating
of tail and double-pigtail stents. J Endourol 14: 195–202,
reduces encrustation of ureteral stents: a preliminary report.
Int J Antimicrob Agents 19: 507–510, 2002.
42. Liatsikos EN, Hom D, Dinlenc CZ, et al: Tail stent ver-
34. Watterson JD, Cadieux PA, Beiko DT, et al: Oxalate-
sus reentry tube: a randomized comparison after percutaneous
degrading enzymes from Oxalobacter formigenes: a novel de-
stone extraction. Urology 59: 15–19, 2002.
vice coating to reduce urinary tract biomaterial-related en-
43. Kay A: Boston Scientific/Microvasive’s Polaris™ dual
crustation. J Endourol 17: 269 –274, 2003.
durometer ureteral stent: the next generation in stent materi-
35. Multanen M, Tammela TL, Laurila M, et al: Biocompat-
als and design (abstract). J Endourol 16(suppl): 8, 2002.
ibility, encrustation and biodegradation of ofloxacin and silver
44. Mardis HK: Standard specification for ureteral stents: a
nitrate coated poly-L-lactic acid stents in rabbit urethra. Urol
comparative evaluation of newer devices (abstract). J Urol
Res 30: 227–232, 2002. 165(suppl): 362, 2001.
36. Phipps JH, and Tyrrell NJ: Transilluminating ureteric
45. Macaluso JN Jr, Deutsch JS, Goodman JR, et al: The use
stents for preventing operative ureteric damage. Br J Obstet
of the Magnetip double-J ureteral stent in urological practice.
Gynaecol 99: 81–86, 1992.
J Urol 142: 701–703, 1989.
37. Tschada RK, Henkel TO, Junemann KP, et al: Spiral-
46. Mykulak DJ, Herskowitz M, and Glassberg KI: Use of
reinforced ureteral stent: an alternative for internal urinary diversion. J Endourol 8: 119 –123, 1994.
magnetic internal ureteral stents in pediatric urology: retrieval
38. Stoller ML, Schwartz BF, Frigstad JR, et al: An in vitro
without routine requirement for cystoscopy and general anes-
assessment of the flow characteristics of spiral-ridged and
thesia. J Urol 152: 976 –977, 1994.
smooth-walled JJ ureteric stents. BJU Int 85: 628 –631, 2000.
47. Taylor WN, and McDougall IT: Minimally invasive ure-
39. Olweny EO, Portis AJ, Sundaram CP, et al: Evaluation
teral stent retrieval. J Urol 168: 2020 –2023, 2002.
of a chronic indwelling prototype mesh ureteral stent in a
48. Antimisiaris SG, Siablis D, Liatsikos E, et al: Liposome-
porcine model. Urology 56: 857–862, 2000.
coated metal stents: an in vitro evaluation of controlled-re-
40. Brewer AV, Elbahnasy AM, Bercowsky E, et al: Mecha-
lease modality in the ureter. J Endourol 14: 743–747, 1999.
nism of ureteral stent flow: a comparative in vivo study. J
49. Slepian MJ: Polymeric endoluminal gel paving: thera-
Endourol 13: 269 –271, 1999.
peutic hydrogel barriers and sustained drug delivery depots
41. Dunn MD, Portis AJ, Kahn SA, et al: Clinical effective-
for local arterial wall biomanipulation. Semin Interven Cardiol
ness of new stent design: randomized single-blind comparison
1: 103–116, 1996.
UROLOGY 64 (1), 2004
VITAMINES ÉNERGÉTIQUES Les facteurs qui influent sur l’état nutritionnel d’une POUR HOMMES 50+ personne étant nombreux, une multivitamine constitue un choix judicieux pour favoriser une bonne santé générale. CUP 0 646420 6561 2 NPN Parmi ces facteurs, on compte un régime déséquilibré qui 80026621 ne comprend pas suffisamment de grains entiers, de fruits et de l�
IMPORTANT LEGAL NOTICE - The information on this site is subject to a disclaimer and a copyright notice. (Free movement of goods - Medicinal products - Importation - Application for marketing authorisation under REFERENCE to the Court under Article 234 EC by the Oberverwaltungsgericht für das Land Nordrhein-Westfalen (Germany) for a preliminary ruling in the proceedings pending before that c