APPLICATION NOTE NanoDrop 2000/2000c
Quality Control of Small Molecule Pharmaceuticals using Spectrophotometry
Introduction Experimental Procedures
The control of pharmaceutical product quality is essential to
Test samples were analyzed per the US Pharmacopeia
ensure safe and effective use in patients. The Thermo
specifications that unknown concentrations of acetaminophen
Scientific NanoDrop 2000 Spectrophotometer offers a rapid
be quantified using absorbance at 244 nm against an
and cost-effective method for spot-checking batch quality at
acetaminophen standard curve prepared from a known
various points in the pharmaceutical production line. The
standard of ≥ 99.9% purity. Prior to experimentation, the
concentrations of many compounds can be determined based
absorbance maximum of acetaminophen was confirmed as
upon UV-Visible absorbance, and, in some instances, purity
can also be monitored by analyzing spectral data.
In this study, UV absorbance spectra measured using a NanoDrop™ 2000 Spectrophotometer were used to compare several acetaminophen samples from various sources with a known standard. In addition, reproducibility was assessed within each sample to determine the affects of aggregates or light scattering in the unknown samples. The auto-ranging pathlength capabilities (fig. 1) of the NanoDrop 2000 allow this instrument to quantify pharmaceutical samples across a much broader concentration range than is possible using a conventional cuvette-based
spectrophotometer. By automatically selecting the optimum
Figure 2: Absorbance spectrum of 6 mg/mL acetaminophen
pathlength (ranging from 1.0 mm to 0.05 mm) the NanoDrop
standard. Absorbance maximum at 244 nm was used for
2000 can accurately measure the absorbance of a sample across
a dynamic range nearly 300 fold greater than that of a cuvette-
Six samples of acetaminophen were compared to the standards.
Each of these samples was dissolved in 100 mL deionized H2O
by incubating for 4 hr at room temperature on an orbital shaker (~100 rpm), resulting in ~5 mg/mL solutions. A 6 mg/mL standard was prepared using 600 mg acetaminophen (Sigma-Aldrich, A7085) dissolved in 100 mL deionized H2O. A standard curve was prepared by performing
a serial dilution of this standard. Absorbance of triplicate, 1 μL aliquots of each dilution were measured at 244 nm using the NanoDrop 2000 UV-Vis application. A linear relationship
Figure 1: Sample liquid column during measurement, showing
between absorbance and concentration was observed
the 1.0 mm (left) and 0.1 mm (right) pathlengths.
throughout the entire standard curve range (fig. 3). Test
acetaminophen solutions were measured against this standard
curve in order to experimentally determine concentrations and
calculate the total acetaminophen content.
Thermo Fisher Scientific – NanoDrop Products Wilmington, Delaware USA Technical support: [email protected] 302-479-7707 www.nanodrop.com
APPLICATION NOTE NanoDrop 2000/2000c
equiv 20 thle pa 10 (1.0mm Concentration (mg/mL) Figure 3: Standard curve of acetaminophen absorbance at 244
Figure 4: Absorbance spectrum of unknown acetaminophen
sample “B.” Nearly identical absorbance spectrum to that observed from measurements of the standard (fig. 2) suggests a
Three separate solutions for each test acetaminophen sample
Conclusion
were measured in triplicate to assess reproducibility (table 1). Reproducibility was very good; CV values of less than 1 %
The linear relationship between acetaminophen concentration
and absorbance at 244 nm makes spectrophotometry an ideal
method to verify batch concentrations. The Thermo Scientific
Replicate Concentration
NanoDrop 2000 Spectrophotometer can be employed to
A244 (n=3)
provide a rapid and accurate spot-check of pharmaceutical
products. The instrument’s auto-ranging pathlength greatly
reduces the need for sample dilution. The full spectrum
display makes this instrument ideal for providing insight into
the purity of various batches of product. Utilizing a traditional
cuvette-based system, a greater number of large-volume serial
dilutions would be required in order to dilute the samples from
their stock concentrations to a measureable concentration
range. By reducing the number of dilutions required, the
NanoDrop 2000 greatly reduces this potential source of error.
In addition, the instrument’s short measurement cycle and
general ease of use greatly increases the rate at which samples
can be processed, making it possible to implement multiple
quality control checks throughout the production process.
Table 1: Acetaminophen content of six unknown samples
References
Absorbance spectra of the test acetaminophen samples (fig. 4) showed a nearly identical profile to the standard solution,
1. United States Pharmacopeia and Natural Formulary (USP 29 NF
24). Supplement No. 2. Rockville, MD: United States
Thermo Fisher Scientific – NanoDrop Products Wilmington, Delaware USA Technical support: [email protected] 302-479-7707 www.nanodrop.com